Skip to main content

Table 4 Values of hemodynamic parameters in normal functioning mechanical prothesis based on valve type and sizes within second follow-up

From: Assessment of normal hemodynamic profile of mechanical pulmonary prosthesis by doppler echocardiography: a prospective cross-sectional study

ValveSubtypes

Size(mm)

Number

PPG (mm Hg)

MPG (mm Hg)

PV (m/sec)

PVAT (ms)

DVI

EOA (cm2)

EOA/BSA (cm2/m2)

St. Jude

21

1

21

13

2.30

85

0.6

2.08

1.22

23

14

23.5 (10–45)

14 (6–30)

2.25 (1.50–3.20)

99.5 (65–140)

0.6 (0.3–0.9)

2.7 (1.25–3.74)

1.7 (0.73–2.91)

25

29

13 (7–73)

8 (2.5–45)

1.70 (1.10–4.10)

82 (35–185)

0.5 (0.3–0.9)

2.45 (1.47–4.42)

1.51 (0.86–3.25)

27

4

10.5 (6–16)

6 (4–9)

1.55 (1.10–1.90)

85.5 (75–110)

0.6 (0.4–0.7)

3.43 (2.29–4.01)

1.97 (1.27–2.21)

29

 

.

.

.

.

.

.

.

31

1

14

8

1.80

113

0.3

2.26

1.26

Carbomedics

21

 

.

.

.

.

.

.

.

23

1

25

17

2.40

101

0.6

2.49

1.52

25

7

20 (10–36)

16 (6–21)

2.20 (1.60–2.80)

121 (92–175)

.70 (0.40–0.80)

3.43 (1.96–3.93)

1.94 (1.03–3.97)

27

1

19

11

2.10

98

0.6

3.43

1.91

29

1

14

7

1.80

100

0.6

3.96

2.83

31

 

.

.

.

.

.

.

.

On-X

21

 

.

.

.

.

.

.

.

23

2

28 (20–36)

17 (12–22)

2.60 (2.20–3.00)

107.5 (96–119)

0.75 (0.7–0.8)

3.11 (2.91–3.32)

2.15 (2.08–2.21)

25

7

19 (10–34)

12 (5–20)

2.10 (1.60–2.90)

85 (79–103)

0.5 (0.3–0.8)

2.94 (2.45–3.91)

1.72 (1.29–2.18)

27

1

6

3

1.20

65

0.3

1.72

0.82

29

 

.

.

.

.

.

.

.

31

 

.

.

.

.

.

.

.

  1. The parameters demonstrated in median and range
  2. PPG peak pressure gradient, MPG mean pressure gradient, PV pulmonary velocity, PVAT pulmonary velocity acceleration time, DVI doppler velocity index, EOA effective orifice area, EOA/BSA effective orifice area division to body surface area Functioning Mechanical Prothesis Based on Valve Type and Sizes within First Follow-up